A question that comes up now and then in our Master Classes is how to figure out the volume of an HPLC column. I’d like to share a couple of rules of thumb that I find very useful for estimating column volume, VM. Figuring out the volume inside the column isn’t much of a challenge if the column is empty. Remember back to your high school geometry class that the volume (V) of a cylinder is

V = Πr2h (1)

So for a 150 x 4.6 mm i.d. column, h (height) = L (length) = 150 mm, and r (radius) = 0.5dc (diameter of the column = 4.6 mm), giving (Π)(150)(4.6/2)2 = 2493 mm2 = 2493 µL ≈ 2.5 mL.

4.6 mm i.d. Columns
However, 2.5 mL is not the volume of a chromatographic column because it is packed with particles. The spherical particles have spaces between them and are also porous, so they have spaces within them. These spaces represent the volume of the mobile phase or column volume that we are really interested in. For most HPLC columns, this volume represents about 60–70% of the volume of an empty column. Let’s see how we can make a shortcut for this. If we multiply any factor of Equation 1 by 60%, we should get the column volume. For the 4.6 mm i.d. column, if we multiply Πr2 by 60%, we get 10 — I always like calculations that involve 5’s and 10’s, because I can do them in my head. So for our 4.6 mm i.d. column we now have:

VM ≈ 10 L (2)

   where L is the length in mm and VM is in µL. Usually we are more interested in VM in mL, so we divide by 1000 or Equation (3) can be restated as 

VM ≈ 0.01 L (3)

   now 1% of 150 = 1.5 mL. We can double-check by multiplying the empty volume, 2.5 mL x 60% = 1.5 mL.

Other Column Diameters

Equation (3) is very handy for estimating the volume of a 4.6 mm i.d. column but what do you do if the column is a different diameter? We need a rule of thumb that includes both L and dc. Notice that 0.6 π (dc/2)2 = (0.6 π/2) dc2 = 0.47 dc2. This is close to 0.5 dc2

VM ≈ 0.5 L dc2 (4)

   where L and dc are in mm and VM is in µL. We can use this to figure out the volume of a 50 x 2.1 mm i.d. column that might be used in an LC–MS application: VM = 0.5 x 50 x 2.12 = 110 µL or ≈ 0.1 mL.

2.1 mm i.d. – A Special Case
We can make one more shortcut for the 2.1 mm i.d. column, which is the most popular column diameter other than 4.6 mm. We could use Equation 2 or 3 and adjust for the change in cross-sectional area and get the same result. Since the cross-sectional area is proportional to the ratio of the diameters squared, we get (4.6/2.1)2 = 4.8. We’re talking about estimates here, so we can round 4.8 to 5 for easy mental math. This means that the volume of a 4.6 mm i.d. column is about five times as large as the same column in a 2.1 mm i.d. format. Let’s do a quick test of this. A 50 x 4.6 mm column would have a volume VM ≈ 0.01 x 50 = 0.5 mL (Equation 3). A 2.1 mm i.d. version of this column should have 1/5 of the volume, so 0.5 mL/5 = 0.1 mL = 100 µL. This is close enough to the value we calculated from Equation 4 above.

   So now we have a few simple shortcuts to estimate column volume. One for the 4.6 mm i.d. columns (Equation 2 or 3), one for any column (Equation 4), and one specifically for the 2.1 mm i.d. column (Equation 2 or 3 divided by 5). These estimates should be within about 10% or so of the value you measure chromatographically.

This blog article series is produced in collaboration with John Dolan, best known as one of the world’s foremost HPLC troubleshooting authorities. He is also known for his research with Lloyd Snyder, which resulted in more than 100 technical publications and three books. If you have any questions about this article send them to TechTips@sepscience.com